Apex

Aerosol particle counters are used to determine the air quality by counting and sizing the number of particles in the air. A common controlled environment aerosol particle counters are used in is a cleanroom. Cleanrooms are used extensively in semiconductor device fabrication, biotechnology, pharmaceutics, disk drives, aerospace and other fields that are very sensitive to environmental contamination.

Air Borne Particle Counter


Specification HANDHELD
2016 3016 5016
Range 0.2 - 2.0 µm 0.3 - 25.0 µm (IAQ : 2.5µm) 0.5 - 25.0 µm
Flow Rate 0.1 CFM (2.83 LPM)
Channel 6 Channels
Display Size Touch Screen Display 3.5" (8.89 cm), color TFT
Battery Li-Ion, removable & rechargeable
Data Storage 3,000 sample records, rotating buffer
Concentration Limit 8,000,000 particles / ft3 @ 10% coincidence loss
Communication RS-232 via RJ-45 to a PC or an optional printer
Enclosure High impact injection molded plastic
Specification Solair
110LD 3100 5100 3350 BOULDER COUNTER
Range 0.10 - 1.0 µm 0.3 - 25.0 µm 0.5 - 25.0 µm 0.3 - 25.0 µm 5.0 - 100.0 µm
Flow Rate 1 CFM (2.83 LPM) 100 LPM 28.3 LPM
Channel 6 Channels
Display Size 5.7-inch Color TFT Touch Screen Display
Battery Optional Li-Ion, removable & rechargeable (typical recharge time 7.5 hours)
Data Storage 3000-record Data Storage Memory
Concentration Limit 500,000 particles/ft3 (17,150,000 / m3) @ 5% coincidence loss" 1,000,000 particles/ft3 (35,300,000/m3) @ 10% coincidence loss Suitable for ISO classes 1-8 cleanrooms 400,000 particles/ft3 (14,120,000/m3) @ 10% coincidence loss Suitable for ISO classes 1-8 cleanrooms 500,000 particles/ft3 @ 5% coincidence loss
Communication USB, Ethernet, RS-485, Secure Data Transfer to and Storage on USB Flash Drive
Enclosure Stainless Steel
Specification Apex Portable Apex Z
P3 P5 Z3 Z50
Range 0.3 - 5.0 µm 0.5 - 10.0 µm 0.3 10.0 µm 0.5 - 10.0 µm
Flow Rate 1 CFM (28.3 LPM) 1.0 CFM (28.3 LPM) 100 LPM
Channel 6 Channels 6 Channels
Display Size 5.7 inch color touch screen 7.0 inch (8.89 cm), 1200 x 800 IPS color touch screen
Battery 4 hrs nominal, 3 hrs continuous 2x Li-Ion battery slot capacity, removable and rechargeable, 11 hours nominal use, 50% duty cycle no printing
Data Storage 3,000 records 10,000 Sample Records, 2500 Locations, 250 Groups, 500 Presets, 500 Sample Plans, 500 Certifications, 100 Reports, 250 Users
Concentration Limit 500,000 particles/ft3 (17,150,000/m3 ) @ 5% coincidence loss suitable for ISO classes 1-‐8 cleanrooms 1,000,000 particles / ft3 @ 10% coincidence loss 300,000 particles / ft3 @ 10% coincidence loss
Communication Ethernet or Serial Ethernet, USB standard, Wi-Fi (802.11 b/g/n)
Enclosure 304 stainless steel Chemically Resistant Polycarbonate blend, conductive lining, VHP compatible
Specification Apex Remote Apex Remote With PUMP Remote With PUMP
R3 R5 R02,R03,R05 R3p R5p R02p,R03p,R05p 5012PN 5102PN
Range 0.3 - 5.0 µm 0.5 - 10.0 µm 0.2 - 10.0 µm 0.3 - 5.0 µm 0.5 - 10.0 µm 0.2 - 10.0 µm 0.5 - 25.0 µm 0.2 - 25.0 µm
Flow Rate 1.0 CFM (2.83 LPM) 0.1 CFM 1.0 CFM (2.83 LPM) 0.1 CFM (2.83 LPM) 0.1 CFM (2.83 LPM) 1.0 CFM (2.83 LPM)
Channel 6 Channels 2 Channels
Display Size OLED Nil (LED Indicators for Power, Service)
Battery -
Data Storage -
Concentration Limit 1,000,000 particles/ft3 @10% (34,300,000/m3 ) coincidence loss suitable for ISO classes 1-8 cleanrooms 4,000,000 particles/ft3 @10% (137,200,000/m3 ) coincidence loss suitable for ISO classes 1-8 cleanrooms 1,000,000 particles/ft3 @10% (34,300,000/m3 ) coincidence loss suitable for ISO classes 1-8 cleanrooms 4,000,000 particles/ft3 @10% (137,200,000/m3 ) coincidence loss suitable for ISO classes 1-8 cleanrooms 4,000,000 particles /ft3 @ 10% coinsidence loss 1,000,000 particles /ft3 @ 10% coinsidence loss
Communication 4-20mA communication output 4-20mA communication output
Enclosure 304 Stainless Steel NEMA style 316L stainless steel or powder coated steel
Liquid Particle COunter

Liquid particle counters are used to determine the quality of the liquid passing through them. The size and number of particles can determine if the liquid is clean enough to be used for the designed application. Liquid particle counters can be used to test the quality of drinking water or cleaning solutions, or the cleanliness of power generation equipment, manufacturing parts, or injectable drugs.

Liquid Particle Counter


Specification Nanocount Batch Smapler Remote
30 50 65C LS20 LS60 LPC 0.1 LPC 0.2 LPC 0.3 LPC 0.5 LPC 1.0
Range 30 - 100 nm 50 - 200 nm 60 - 200nm. User defineable 1.0 - 50.0 µm 0.1 - 0.5 µm; 0.2 - 2.0 µm; 0.3 - 3.0 µm; 0.5 - 100.0 µm 0.1 - 0.5 µm 0.2 - 2.0 µm 0.3 - 3.0 µm 0.5 - 100.0 µm 1 - 50 µm; 1 -200 µm
Flow rate 80 mL / minute 100 mL / minute 20 ml/minute 60 ml/min 100 mL / minute. 50 ml/minute; 30 ml/minute
Display Size TFT color touch screen "Nil (LED Indicators for Power, leak alarm Nil "Nil [LED Indicators for Power, Sampling (4-20mA model); Power, Service, Sampling (485 modbus model)]. LED Indicators for Power, Service, Sampling
Battery -
Data Storage Nil Rotating buffer, 2,000 records> Nil Rotating buffer 2,000 records (modbus models) Rotating buffer, 1,000 records
Concentration Limit 1,000,000 counts / ml @ 5% coinsidence error 200,000 counts / ml @ 5% coinsidence error 64,000 counts/ml @ 5% coincidence error. 4,000 counts/ml @ 5% coincidence error. 40,000 P/mL; 50,000 P/mL
Communication Ethernet Modbus TCP/IP, RS-232 via RJ-45 to PC, RS-485 Modbus Industrial RJ45 for RS-485, Ethernet; sealed DB9 for 4-20mA RS-485 MODBUS RS232/485 via RJ45 to PC, RS485 modbus, ASCII, Modbus or MR protocol. RS485/modbus via RJ-45 to PC, 4-20mA or ASCII
Enclosure Stainless Steel 304L Stainless Steel Polypropylene, white Stainless Steel Stainless Steel
Micorbial Air Sampler

Lighthouse makes counting of Viables user friendly with a touchscreen UI and accuracy in both mobile and remote sampling systems. Proven microbial sampler technology for high-end pharma, medical and hospital applications.

Microbial Air Sampler


Specification Active Count AC Remote
100 100H 10025H R RP
Flow Rate 316L Stainless Steel with 300 holes, 19m/sec, d50: 1µm 100 liters/minute ± 4%3 25 liters/minute ± 5%3 1 CFM.
Impactor Head Aluminum head with 300 holes, 19 m/sec (stainless steel option available) 316L Stainless Steel with 300 holes, 19m/sec, d50: 1µm 316L Stainless Steel with 88 holes, 19m/sec, d50: 1µm
Sample Presets 8 sample size presets from 20L - 2000L
Display Size 3.5 inch (8.9 cm) color touch screen with adjustable brightness LED Indicator for Power, Flow, Sampling.
Battery 6 Hours continuous sampling or 8 hours normal sampling use1 , Li-ion rechargeable, 4 hrs charge time 8 Hrs. continuous, 10 Hrs. normal sampling2, Li-ion rechargeable, 4 Hrs. charge time
Data Storage Saved on USB flash drive, .csv file format Unlimited; saved on USB flash drive in .csv file format RS-485 MODBUS
Communication USB 2.0 USB 2.0 - 8GB Flash Drive included RS-485 MODBUS
Enclosure 316L Stainless Steel Stainless Steel
Oil Particle Counter

Oil Particle Counter


It can be used for oil pollution degree detection , includes hydraulic oil, aviation kerosene, lubricant oil, transformer oil( insulation oil ), steam turbine oil, gear oil, engine oil, and water-based hydraulic oil,etc.

How can we help you?

If you have any questions, comments, or feedback, please feel free to contact us. We‘re ready to help!

Contact US